Introduction
- 3 mm Allen keyPhillips screwdriver, size 1Knife with a smooth, flat blade
These steps are not difficult but need to be followed carefully.
For more information on replacing the fascia filter, please refer to page 45 in the service manual.
What you need
Step 1
Fascia Film
- Open the device
- 1.Place the device on a non-slip surface with the top facing downwards
- 2.Remove the filter cassette from the top part of the housing.
- 3.Undo and remove the 6 screws (34)
Open the device
1.Place the device on a non-slip surface with the top facing downwards
2.Remove the filter cassette from the top part of the housing.
3.Undo and remove the 6 screws (34)
1024
Step 2
- Remove the mains supply unit
- Disconnect the power board/mains supply unit connecting cable (45) from the power board.
- Slightly lift mains supply unit (44) and disconnect the mains input connector from the mains supply unit (44).
- Lift the mains supply unit out of the top part of the housing.
- Disconnect the power board/mains supply unit connecting cable from the mains supply unit.
Remove the mains supply unit
Disconnect the power board/mains supply unit connecting cable (45) from the power board.
Slightly lift mains supply unit (44) and disconnect the mains input connector from the mains supply unit (44).
Lift the mains supply unit out of the top part of the housing.
Disconnect the power board/mains supply unit connecting cable from the mains supply unit.
Step 3
- Remove the power board. Weinmann VENTImotion and VENTIlogic Power Board Replacement
Remove the power board. Weinmann VENTImotion and VENTIlogic Power Board Replacement
Step 4
- Remove box and filter holder
- Release the pressure-side hose (74) from the box.
- Lift box (58) and filter holder (37) out of the top part of the housing together.
- Disconnect the box and the filter holder from one another by undoing the intake-side hose (73) from both parts.
Remove box and filter holder
Release the pressure-side hose (74) from the box.
Lift box (58) and filter holder (37) out of the top part of the housing together.
Disconnect the box and the filter holder from one another by undoing the intake-side hose (73) from both parts.
Step 5
- Remove the control board and the display Weinmann VENTImotion and VENTIlogic Control Board Replacement
- Undo fascia film (33) using a knife-blade and carefully peel it off.
- Use a little 70% isopropanol to degrease this part of the housing.
Remove the control board and the display Weinmann VENTImotion and VENTIlogic Control Board Replacement
Undo fascia film (33) using a knife-blade and carefully peel it off.
Use a little 70% isopropanol to degrease this part of the housing.
Step 6
- Peel the protective film off the inside of the new fascia film and affix fascia film (33) to this area of the housing.
- Ensure that the ribbon cable is pushed cleanly through the opening in the housing without any kinks.
Peel the protective film off the inside of the new fascia film and affix fascia film (33) to this area of the housing.
Ensure that the ribbon cable is pushed cleanly through the opening in the housing without any kinks.
Step 7
- Carefully pull the outer protective film off the new fascia film using a fingernail.
Carefully pull the outer protective film off the new fascia film using a fingernail.
Step 8
- Refit the display and the control board Weinmann VENTImotion and VENTIlogic Control Board Replacement
Refit the display and the control board Weinmann VENTImotion and VENTIlogic Control Board Replacement
Step 9
- Refit the box and filter holder.
- Connect the box and the filter holder to one another by reconnecting the intake side hose (73) from both parts.
- Refit box (58) and filter holder (37) back on the top part of the housing together.
- Connect the pressure-side hose (74) from the box.
Refit the box and filter holder.
Connect the box and the filter holder to one another by reconnecting the intake side hose (73) from both parts.
Refit box (58) and filter holder (37) back on the top part of the housing together.
Connect the pressure-side hose (74) from the box.
Step 10
- Refit the power board Weinmann VENTImotion and VENTIlogic Power Board Replacement
Refit the power board Weinmann VENTImotion and VENTIlogic Power Board Replacement
Step 11
- Refit the mains supply unit
- Reconnect the power board/mains supply unit connecting cable from the mains supply unit.
- Put the mains supply unit back in the top part of the housing.
- Reconnect the mains input connector from the mains supply unit (44).
- Reconnect the power board/mains supply unit connecting cable (45) from the power board.
Refit the mains supply unit
Reconnect the power board/mains supply unit connecting cable from the mains supply unit.
Put the mains supply unit back in the top part of the housing.
Reconnect the mains input connector from the mains supply unit (44).
Reconnect the power board/mains supply unit connecting cable (45) from the power board.
Step 12
- Close the device
- Hold the bottom part of the housing (35) up to the side of the top part of the housing (36)
- Plug the connecting cables for the humidifier (42) and the alarm (43) onto the relevant connectors.
- Check that all hoses and cables are plugged on firmly.
- Place the bottom part of the housing (35) on the top part of the housing (36). Ensure that no cables or hoses are trapped or bent.
- Now screw the top part of the housing tight using the 6 screws (34).
- Then turn the device back over
Close the device
Hold the bottom part of the housing (35) up to the side of the top part of the housing (36)
Plug the connecting cables for the humidifier (42) and the alarm (43) onto the relevant connectors.
Check that all hoses and cables are plugged on firmly.
Place the bottom part of the housing (35) on the top part of the housing (36). Ensure that no cables or hoses are trapped or bent.
Now screw the top part of the housing tight using the 6 screws (34).
Then turn the device back over
Step 13
- Test the device
- For detailed information see page 13 of VENTImotion & VENTIlogic Service and Repair Instructions
Test the device
For detailed information see page 13 of VENTImotion & VENTIlogic Service and Repair Instructions
To reassemble your device, follow these instructions in reverse order.
Cancel: I did not complete this guide.
2 other people completed this guide.
Author
with 2 other contributors
Jay Vidinoff
Member since: 04/07/2020
610 Reputation
2 Guides authored
Badges:
5
+2 more badges
Team
Cal Poly, Team S7-G19, Paton Spring 2020
Member of Cal Poly, Team S7-G19, Paton Spring 2020
CPSU-PATON-S20S7G19
3 Members
4 Guides authored